AssayIndex
Glossary

Terms beginning with G

8 of 238 glossary terms.

Gradient
Programmed change in mobile-phase composition during a run. Gradient duration determines resolving power and is therefore part of the result, not metadata.
Gradient slope
Percentage points of organic solvent per minute. Lower slope, better resolution, longer run.
see also Gradient
Guard column
A short disposable column before the analytical column, protecting it from particulates and strongly retained matrix. Shifts retention slightly and must be accounted for.
Ghost peak
A peak arising from material retained on the column from a previous injection and released by a later gradient. Detected by a blank injection.
see also Carry-over
Gravimetric fill check
Weighing a vial before and after removal of contents to determine the actual fill. The only direct measurement of label-claim compliance.
GC headspace
Gas chromatography of the vapour above a heated sealed sample. The standard technique for residual solvents.
Gel-clot
A limit-test LAL format: a firm gel forms or it does not. Gives a pass or fail against a stated sensitivity, never a number.
see also LAL
Gradient normalisation
Converting a retention time measured on one gradient duration to the index reference scale, so retentions from different laboratories can be compared.