AssayIndex

ASY-{CAT}-2288

Retatrutide · 30 mg declared · Asymchem Laboratories (Tianjin) Co., Ltd. · manufactured 26 Aug 2026

CHARACTERISED
Lot code
ASY-{CAT}-2288
Supplier
ASY — Asymchem Laboratories (Tianjin) Co., Ltd.
Compound
Retatrutide
Compound class
Incretin analogues and co-agonists
Declared label claim
30mg
Manufacture date
26 Aug 2026
Index quarter
2026-Q3
Independent assays
11
Laboratories involved
Kestrel, PeptideMeter, Janoshik, VendorInvestigate, Nordanalyt
Provenance mix
Buyer-sourced ×10 · Laboratory-retained ×1
Section 1

Reproducibility across independent tests

Every report below is an observation of the same physical material. What they disagree about is measurement; what they agree about is the lot. The size of the disagreement is compared against what the contributing laboratories' own repeatability figures predict.

Independent assays 11
Median purity 98.45% n=11
Spread (max − min) 1.90pp n=11
Standard deviation 0.595pp n=11
95.596.397.197.998.799.5100.3This lot (ASY-{CAT}-2288)This lot (ASY-{CAT}-2288): Q1 98.25, median 98.45, Q3 98.84, n=11outlier 99.82n=11All RetatrutideAll Retatrutide: Q1 97.96, median 98.37, Q3 98.82, n=492outlier 95.81outlier 96.13outlier 96.29outlier 96.32outlier 96.46outlier 96.48outlier 96.52outlier 96.54outlier 96.54outlier 96.63outlier 96.64n=492 Purity (% area)
Figure 1. This lot's independent purity results against the whole Retatrutide record set in the index. The vertical line is the index acceptance specification for the incretin analogues and co-agonists class, 96.5 % area. A narrow box on the top row against a wide box on the bottom row is the ideal reading: it means the disagreement between lots is much larger than the disagreement within this one.
Section 2

Member reports 11

Every report the index holds on this lot. Each is a page carrying its own chromatogram, mass spectrum, mass balance, provenance and caveats.

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-041783ASY-{CAT}-2288Kestrel2026-Q399.446.46101.4Buyer-sourced0PASS
AX-2026-041784ASY-{CAT}-2288PeptideMeter2026-Q398.156.2691.9Buyer-sourced0PASS
AX-2026-041785ASY-{CAT}-2288Janoshik2026-Q397.965.9290.6Laboratory-retained0PASS
AX-2026-041786ASY-{CAT}-2288VendorInvestigate2026-Q398.786.59Buyer-sourced0PASS
AX-2026-041787ASY-{CAT}-2288VendorInvestigate2026-Q398.896.29103.0Buyer-sourced0PASS
AX-2026-041788ASY-{CAT}-2288Janoshik2026-Q398.455.77Buyer-sourced0PASS
AX-2026-041789ASY-{CAT}-2288Janoshik2026-Q398.376.1999.4Buyer-sourced0PASS
AX-2026-041790ASY-{CAT}-2288Nordanalyt2026-Q397.926.35Buyer-sourced0PASS
AX-2026-041791ASY-{CAT}-2288Janoshik2026-Q398.495.63Buyer-sourced0PASS
AX-2026-041792ASY-{CAT}-2288Kestrel2026-Q399.826.81Buyer-sourced0PASS
AX-2026-041793ASY-{CAT}-2288Nordanalyt2026-Q398.346.16104.8Buyer-sourced0PASS
Section 3

By laboratory

Where more than one laboratory assayed this lot, the difference between them is a paired observation of their relative bias — the same material, the same time window, different instruments. Those pairs are what the index's bias estimates are built from.

LaboratoryReportsPurity results (% area)MedianGradientPublished bias (pp)Bias-adjusted
Kestrel299.44, 99.8299.6315min+0.2499.39
PeptideMeter198.1598.1528min-0.0598.20
Janoshik497.96, 98.45, 98.37, 98.4998.4130min+0.0098.41
VendorInvestigate298.78, 98.8998.8432min-0.1198.95
Nordanalyt297.92, 98.3498.1335min+0.0298.11

This lot contributes 10 paired comparisons to the inter-lab analysis.

Section 4

The supplier's certificate for this lot

One number the supplier wrote, beside the median of the numbers other people measured on the same physical material. That comparison is only available because the index groups by lot.

FigureSupplier COAThis reportIndex note
Purity (% area)98.6599.44divergence +0.20 pp vs batch median
Total impurities (% area)1.350.56
Largest single impurity (% area)0.940.24
Water (%)3.86.46
Counter-ion (%)3.98.35
Peptide content (%w/w)91.784.98
Endotoxin (EU/mg)not statednot determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Impurity fingerprint

The named impurities reported across this lot's assays, with how consistently each was seen. A fingerprint is what allows two lot numbers held by different companies to be identified as one bulk.

Named impurityReported inMedian % areaRange
des-Aib20 analogue11 / 110.340.04–0.60
[D-Aib2]-diastereomer11 / 110.850.08–1.39
truncated (5-39)10 / 110.110.06–0.21
oxidised Trp (+16 Da)10 / 110.040.02–0.08
incomplete Fmoc removal (+222 Da)10 / 110.050.03–0.13
C18 diacid regioisomer4 / 110.030.02–0.03
aspartimide (−18 Da)4 / 110.040.03–0.04
Section 7

What this rollup does not establish