Mazdutide by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for Mazdutide, with a class minimum gradient of 25 min, the method is a live explanation.

33 recordsmedian 98.11 %spec 96.5 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik1330minHigh98.57+0.0098.57ADEQUATE
PeptideMeter7 n=728minHigh97.94-0.0597.99ADEQUATE
Medutest6 n=625minUnit98.36+0.0898.28ADEQUATE
VendorInvestigate3 n=332minHigh97.32-0.1197.43ADEQUATE
Nordanalyt2 n=235minHigh97.61, 97.32+0.02ADEQUATE
Kestrel2 n=215minUnit97.99, 98.11+0.24BELOW CLASS MINIMUM
96.897.397.998.599.099.6100.2JanoshikJanoshik: Q1 98.10, median 98.57, Q3 99.02, n=13n=13PeptideMeterPeptideMeter: Q1 97.91, median 97.94, Q3 98.20, n=7outlier 97.02outlier 98.74n=7MedutestMedutest: Q1 97.98, median 98.36, Q3 98.54, n=6n=6VendorInvestigateVendorInvestigate: Q1 97.30, median 97.32, Q3 97.80, n=3n=3NordanalytNordanalyt: Q1 97.39, median 97.47, Q3 97.54, n=2n=2KestrelKestrel: Q1 98.02, median 98.05, Q3 98.08, n=2n=2 Purity (% area)
Figure. Mazdutide purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.