GHRP-2 by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for GHRP-2, with a class minimum gradient of 15 min, the method is a live explanation.

97 recordsmedian 99.22 %spec 99.0 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik4830minHigh99.19+0.0099.19ADEQUATE
Medutest1425minUnit99.62+0.0899.54ADEQUATE
Nordanalyt1435minHigh99.21+0.0299.19ADEQUATE
PeptideMeter1128minHigh99.32-0.0599.37ADEQUATE
VendorInvestigate6 n=632minHigh98.78-0.1198.89ADEQUATE
Kestrel4 n=415minUnit99.50+0.2499.27ADEQUATE
97.798.198.598.999.399.7100.1JanoshikJanoshik: Q1 98.89, median 99.19, Q3 99.50, n=48outlier 97.91n=48PeptideMeterPeptideMeter: Q1 98.97, median 99.32, Q3 99.56, n=11n=11MedutestMedutest: Q1 99.12, median 99.62, Q3 99.70, n=14n=14VendorInvestigateVendorInvestigate: Q1 98.48, median 98.78, Q3 99.06, n=6n=6NordanalytNordanalyt: Q1 98.95, median 99.21, Q3 99.59, n=14n=14KestrelKestrel: Q1 99.14, median 99.50, Q3 99.86, n=4n=4 Purity (% area)
Figure. GHRP-2 purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.