Section 1

Reproducibility across independent tests

Every report below is an observation of the same physical material. What they disagree about is measurement; what they agree about is the lot. The size of the disagreement is compared against what the contributing laboratories' own repeatability figures predict.

Independent assays 13
Median purity 98.50% n=13
Spread (max − min) 1.08pp n=13
Standard deviation 0.330pp n=13
97.698.098.598.999.399.7100.1This lot (WXT-2402-W3946)This lot (WXT-2402-W3946): Q1 98.28, median 98.50, Q3 98.79, n=13n=13All SemaxAll Semax: Q1 98.65, median 99.17, Q3 99.63, n=160n=160 Purity (% area)
Figure 1. This lot's independent purity results against the whole Semax record set in the index. The vertical line is the index acceptance specification for the neuroactive and cognitive peptides class, 98.5 % area. A narrow box on the top row against a wide box on the bottom row is the ideal reading: it means the disagreement between lots is much larger than the disagreement within this one.
Section 2

Member reports 13

Every report the index holds on this lot. Each is a page carrying its own chromatogram, mass spectrum, mass balance, provenance and caveats.

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2024-005401WXT-2402-W3946Kestrel2024-Q298.096.85Supplier-supplied0FAIL
AX-2024-005402WXT-2402-W3946Kestrel2024-Q298.506.4589.4Laboratory-retained1MARGINAL
AX-2024-005403WXT-2402-W3946Kestrel2024-Q298.796.6193.4Supplier-supplied0PASS
AX-2024-005404WXT-2402-W3946Medutest2024-Q298.206.7792.6Supplier-supplied0FAIL
AX-2024-005405WXT-2402-W3946Medutest2024-Q298.506.59103.2Laboratory-retained1MARGINAL
AX-2024-005406WXT-2402-W3946Medutest2024-Q298.617.19Supplier-supplied1FAIL
AX-2024-005407WXT-2402-W3946Medutest2024-Q298.646.8095.2Supplier-supplied0PASS
AX-2024-005408WXT-2402-W3946Janoshik2024-Q298.996.60Supplier-supplied1MARGINAL
AX-2024-005409WXT-2402-W3946Nordanalyt2024-Q298.346.76Buyer-sourced0FAIL
AX-2024-005410WXT-2402-W3946Medutest2024-Q299.176.77Supplier-supplied0PASS
AX-2024-005411WXT-2402-W3946Medutest2024-Q298.286.72Buyer-sourced0FAIL
AX-2024-005412WXT-2402-W3946Nordanalyt2024-Q298.846.52105.8Supplier-supplied0PASS
AX-2024-005413WXT-2402-W3946Medutest2024-Q298.206.68Supplier-supplied0FAIL
Section 3

By laboratory

Where more than one laboratory assayed this lot, the difference between them is a paired observation of their relative bias — the same material, the same time window, different instruments. Those pairs are what the index's bias estimates are built from.

LaboratoryReportsPurity results (% area)MedianGradientPublished bias (pp)Bias-adjusted
Kestrel398.09, 98.50, 98.7998.5015min+0.2498.26
Medutest798.20, 98.50, 98.61, 98.64, 99.17, 98.28, 98.2098.5025min+0.0898.42
Janoshik198.9998.9930min+0.0098.99
Nordanalyt298.34, 98.8498.5935min+0.0298.57

This lot contributes 6 paired comparisons to the inter-lab analysis.

Section 4

The supplier's certificate for this lot

One number the supplier wrote, beside the median of the numbers other people measured on the same physical material. That comparison is only available because the index groups by lot.

FigureSupplier COAThis reportIndex note
Purity (% area)98.7398.09divergence +0.23 pp vs batch median
Total impurities (% area)not stated1.91
Largest single impurity (% area)not stated1.26
Water (%)4.56.85
Counter-ion (%)not stated6.75
Peptide content (%w/w)90.7not determined
Endotoxin (EU/mg)not statednot determined
Method stated on certificateNOno method section — the result cannot be reconciled against any other
Section 5

Impurity fingerprint

The named impurities reported across this lot's assays, with how consistently each was seen. A fingerprint is what allows two lot numbers held by different companies to be identified as one bulk.

Named impurityReported inMedian % areaRange
diketopiperazine (−18 Da)13 / 130.900.45–1.26
N-acetyl semax (+42 Da)13 / 130.220.12–0.31
des-Met (−131 Da)13 / 130.170.11–0.35
Met1 sulfoxide (+16 Da)13 / 130.140.10–0.21
Section 7