AssayIndex

SSA-2604-L8146

Semax · 5 mg declared · Shanghai Sigma-Audley New Material Technology Co., Ltd. · manufactured 17 Apr 2026

CHARACTERISED
Lot code
SSA-2604-L8146
Supplier
SSA — Shanghai Sigma-Audley New Material Technology Co., Ltd.
Compound
Semax
Compound class
Neuroactive and cognitive peptides
Declared label claim
5mg
Manufacture date
17 Apr 2026
Index quarter
2026-Q2
Independent assays
14
Laboratories involved
Medutest, Janoshik, Nordanalyt
Provenance mix
Buyer-sourced ×12 · Supplier-supplied ×2
Section 1

Reproducibility across independent tests

Every report below is an observation of the same physical material. What they disagree about is measurement; what they agree about is the lot. The size of the disagreement is compared against what the contributing laboratories' own repeatability figures predict.

Independent assays 14
Median purity 98.60% n=14
Spread (max − min) 0.83pp n=14
Standard deviation 0.257pp n=14
97.998.298.699.099.399.7100.1This lot (SSA-2604-L8146)This lot (SSA-2604-L8146): Q1 98.51, median 98.59, Q3 98.88, n=14n=14All SemaxAll Semax: Q1 98.83, median 99.26, Q3 99.63, n=177n=177 Purity (% area)
Figure 1. This lot's independent purity results against the whole Semax record set in the index. The vertical line is the index acceptance specification for the neuroactive and cognitive peptides class, 98.5 % area. A narrow box on the top row against a wide box on the bottom row is the ideal reading: it means the disagreement between lots is much larger than the disagreement within this one.
Section 2

Member reports 14

Every report the index holds on this lot. Each is a page carrying its own chromatogram, mass spectrum, mass balance, provenance and caveats.

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-043222SSA-2604-L8146Medutest2026-Q298.568.6695.8Buyer-sourced0PASS
AX-2026-043223SSA-2604-L8146Janoshik2026-Q298.628.0091.5Buyer-sourced0PASS
AX-2026-043224SSA-2604-L8146Nordanalyt2026-Q298.508.9798.3Buyer-sourced0PASS
AX-2026-043225SSA-2604-L8146Medutest2026-Q298.868.60Buyer-sourced0PASS
AX-2026-043226SSA-2604-L8146Medutest2026-Q299.038.6091.3Buyer-sourced0PASS
AX-2026-043227SSA-2604-L8146Janoshik2026-Q298.488.49Buyer-sourced0FAIL
AX-2026-043228SSA-2604-L8146Janoshik2026-Q298.259.22Buyer-sourced0FAIL
AX-2026-043229SSA-2604-L8146Janoshik2026-Q298.578.3797.8Buyer-sourced0PASS
AX-2026-043230SSA-2604-L8146Medutest2026-Q298.898.5990.3Buyer-sourced0PASS
AX-2026-043231SSA-2604-L8146Janoshik2026-Q299.088.3996.0Supplier-supplied0PASS
AX-2026-043232SSA-2604-L8146Medutest2026-Q298.548.09Buyer-sourced1FAIL
AX-2026-043233SSA-2604-L8146Janoshik2026-Q298.349.02103.1Buyer-sourced0FAIL
AX-2026-043234SSA-2604-L8146Janoshik2026-Q298.868.44106.6Buyer-sourced0PASS
AX-2026-043235SSA-2604-L8146Janoshik2026-Q298.898.3898.1Supplier-supplied0PASS
Section 3

By laboratory

Where more than one laboratory assayed this lot, the difference between them is a paired observation of their relative bias — the same material, the same time window, different instruments. Those pairs are what the index's bias estimates are built from.

LaboratoryReportsPurity results (% area)MedianGradientPublished bias (pp)Bias-adjusted
Medutest598.56, 98.86, 99.03, 98.89, 98.5498.8625min+0.0898.78
Janoshik898.62, 98.48, 98.25, 98.57, 99.08, 98.34, 98.86, 98.8998.5930min+0.0098.59
Nordanalyt198.5098.5035min+0.0298.48

This lot contributes 3 paired comparisons to the inter-lab analysis.

Section 4

The supplier's certificate for this lot

One number the supplier wrote, beside the median of the numbers other people measured on the same physical material. That comparison is only available because the index groups by lot.

FigureSupplier COAThis reportIndex note
Purity (% area)99.1598.56divergence +0.55 pp vs batch median
Total impurities (% area)0.851.44
Largest single impurity (% area)0.590.60
Water (%)not stated8.66
Counter-ion (%)not stated5.94
Peptide content (%w/w)91.785.15
Endotoxin (EU/mg)0.40not determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Impurity fingerprint

The named impurities reported across this lot's assays, with how consistently each was seen. A fingerprint is what allows two lot numbers held by different companies to be identified as one bulk.

Named impurityReported inMedian % areaRange
N-acetyl semax (+42 Da)14 / 140.340.15–0.73
Met1 sulfoxide (+16 Da)14 / 140.240.07–0.64
des-Met (−131 Da)14 / 140.150.10–0.19
diketopiperazine (−18 Da)14 / 140.400.10–1.07
Section 7

What this rollup does not establish