AssayIndex

CPC-2504-F2176

Selank · 20 mg declared · Chinese Peptide Company Co., Ltd. · manufactured 16 Apr 2025

CHARACTERISED
Lot code
CPC-2504-F2176
Supplier
CPC — Chinese Peptide Company Co., Ltd.
Compound
Selank
Compound class
Neuroactive and cognitive peptides
Declared label claim
20mg
Manufacture date
16 Apr 2025
Index quarter
2025-Q2
Independent assays
14
Laboratories involved
PeptideMeter, Medutest, Janoshik, Kestrel, VendorInvestigate
Provenance mix
Buyer-sourced ×13 · Supplier-supplied ×1
Section 1

Reproducibility across independent tests

Every report below is an observation of the same physical material. What they disagree about is measurement; what they agree about is the lot. The size of the disagreement is compared against what the contributing laboratories' own repeatability figures predict.

Independent assays 14
Median purity 99.04% n=14
Spread (max − min) 1.23pp n=14
Standard deviation 0.309pp n=14
97.898.298.698.999.399.7100.1This lot (CPC-2504-F2176)This lot (CPC-2504-F2176): Q1 98.92, median 99.04, Q3 99.24, n=14n=14All SelankAll Selank: Q1 99.15, median 99.48, Q3 99.78, n=139outlier 97.94outlier 97.99outlier 98.07n=139 Purity (% area)
Figure 1. This lot's independent purity results against the whole Selank record set in the index. The vertical line is the index acceptance specification for the neuroactive and cognitive peptides class, 99.0 % area. A narrow box on the top row against a wide box on the bottom row is the ideal reading: it means the disagreement between lots is much larger than the disagreement within this one.
Section 2

Member reports 14

Every report the index holds on this lot. Each is a page carrying its own chromatogram, mass spectrum, mass balance, provenance and caveats.

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2025-020970CPC-2504-F2176PeptideMeter2025-Q298.979.77Buyer-sourced0FAIL
AX-2025-020971CPC-2504-F2176Medutest2025-Q298.898.75Buyer-sourced0FAIL
AX-2025-020972CPC-2504-F2176Janoshik2025-Q298.908.47Buyer-sourced0FAIL
AX-2025-020973CPC-2504-F2176Janoshik2025-Q299.079.42Buyer-sourced0PASS
AX-2025-020974CPC-2504-F2176Kestrel2025-Q298.638.99105.7Buyer-sourced0FAIL
AX-2025-020975CPC-2504-F2176Janoshik2025-Q299.698.94Supplier-supplied1MARGINAL
AX-2025-020976CPC-2504-F2176Janoshik2025-Q299.088.99Buyer-sourced0PASS
AX-2025-020977CPC-2504-F2176VendorInvestigate2025-Q299.019.53Buyer-sourced0PASS
AX-2025-020978CPC-2504-F2176Kestrel2025-Q299.249.38100.4Buyer-sourced0PASS
AX-2025-020979CPC-2504-F2176VendorInvestigate2025-Q298.469.2297.8Buyer-sourced0FAIL
AX-2025-020980CPC-2504-F2176Janoshik2025-Q299.398.57Buyer-sourced0PASS
AX-2025-020981CPC-2504-F2176Janoshik2025-Q299.239.06Buyer-sourced0PASS
AX-2025-020982CPC-2504-F2176PeptideMeter2025-Q299.029.6393.7Buyer-sourced0PASS
AX-2025-020983CPC-2504-F2176PeptideMeter2025-Q299.319.7590.8Buyer-sourced0PASS
Section 3

By laboratory

Where more than one laboratory assayed this lot, the difference between them is a paired observation of their relative bias — the same material, the same time window, different instruments. Those pairs are what the index's bias estimates are built from.

LaboratoryReportsPurity results (% area)MedianGradientPublished bias (pp)Bias-adjusted
PeptideMeter398.97, 99.02, 99.3199.0228min-0.0599.07
Medutest198.8998.8925min+0.0898.81
Janoshik698.90, 99.07, 99.69, 99.08, 99.39, 99.2399.1630min+0.0099.16
Kestrel298.63, 99.2498.9415min+0.2498.70
VendorInvestigate299.01, 98.4698.7332min-0.1198.84

This lot contributes 10 paired comparisons to the inter-lab analysis.

Section 4

The supplier's certificate for this lot

One number the supplier wrote, beside the median of the numbers other people measured on the same physical material. That comparison is only available because the index groups by lot.

FigureSupplier COAThis reportIndex note
Purity (% area)98.5998.97divergence -0.45 pp vs batch median
Total impurities (% area)not stated1.03
Largest single impurity (% area)not stated0.47
Water (%)not stated9.77
Counter-ion (%)not stated8.42
Peptide content (%w/w)91.7not determined
Endotoxin (EU/mg)not statednot determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Impurity fingerprint

The named impurities reported across this lot's assays, with how consistently each was seen. A fingerprint is what allows two lot numbers held by different companies to be identified as one bulk.

Named impurityReported inMedian % areaRange
diketopiperazine (−18 Da)14 / 140.390.12–0.63
tuftsin fragment (−417 Da)14 / 140.210.05–0.39
des-Thr (−101 Da)14 / 140.170.09–0.56
acetyl adduct (+42 Da)14 / 140.110.03–0.16
Section 7

What this rollup does not establish