AssayIndex
Chromatographic purity

Purity by reverse-phase HPLC (% area)

The index reference method RG-2, what it measures, what it does not, and why a purity figure without a stated gradient is uninterpretable.

method reference4 sections3 pharmacopoeial references

Chromatographic purity is the area of the main peak expressed as a percentage of the total integrated area of all peaks in the chromatogram, excluding the injection disturbance and the solvent front. It is a relative quantity. It answers the question "of everything that eluted and absorbed at this wavelength, what fraction was the main peak" and it answers no other question. In particular it does not say how much peptide is in the vial, and it does not say what the main peak is.

Index reference method RG-2

ParameterSetting
ColumnC18, 2.6 µm core–shell, 100 × 4.6 mm, 100 Å
Mobile phase AWater + 0.1 % (v/v) trifluoroacetic acid
Mobile phase BAcetonitrile + 0.1 % (v/v) trifluoroacetic acid
Gradient2 % B hold 2 min → 5 % B at 3 min → 65 % B at 30 min → 95 % B at 32 min, hold 3 min
Flow1.0 mL/min
Column temperature40 °C ± 1 °C
DetectionUV 214 nm, 4 nm bandwidth; secondary channel 280 nm where aromatic residues are present
Injection10 µL of a 0.5 mg/mL solution in mobile phase A
Run time35 min including re-equilibration

The two-minute initial hold at 2 % B exists for one reason: compounds below about 500 Da elute within three minutes of injection, and without the hold their main peak partially merges with the solvent disturbance. Several reports submitted to the index have integrated solvent into a tripeptide main peak. The hold is not optional.

What the gradient length decides

Resolution between two peaks scales with the time the gradient spends passing through their elution window. Halving the gradient duration roughly halves the separation between neighbouring species — and the species that matter most on the acylated incretins are mass-identical diastereomers separated by 0.2 min on a 30-minute method. On a 15-minute method they are one peak.

The consequence is arithmetic rather than opinion: a laboratory running a 15-minute gradient will report a higher purity than a laboratory running a 30-minute gradient on the same vial, because it is integrating more material into the main peak. The index measures the offset at +0.24 pp overall and +0.51 pp on the incretin class. Both figures are on the inter-lab agreement page with their confidence intervals.

Acceptance criteria applied by the index

CriterionRequirementIf not met
Gradient duration≥ 25 min for the incretin class, ≥ 22 min for melanocortins, ≥ 15 min otherwiseStatus set to METHOD LIMITED
Main-peak asymmetry (10 % height)As ≤ 1.50Rule AX-R12; purity marked integration-limited
Resolution to nearest named impurityR ≥ 1.20Rule AX-R12
Plate count (main peak)N ≥ 20,000Method note recorded
Retention within compound windowstated window after gradient normalisationRule AX-R11
Baselineno drift exceeding 2 % of main-peak height across the peakMethod note recorded

Worked example

Integrated areas (µV·s), 214 nm:

  main peak            4 812 660
  impurity A (−0.35)      22 140
  impurity B (−0.68)      13 405
  impurity C (+0.93)       7 890
  impurity D (+1.62)       4 215
  ──────────────────────────────
  total                4 860 310

purity  = 4 812 660 / 4 860 310 × 100 = 99.019 %  → 99.02 %
total impurities                       = 0.98 %
largest single impurity  = 22 140 / 4 860 310 × 100 = 0.46 %

check:  99.02 + 0.98 = 100.00   ✓ closes

Every record in this index closes that check to within 0.02 pp, and the build asserts it. A supplier certificate that does not close triggers rule AX-R02.

Normative references

Cited because they are the documents the acceptance criteria above are taken from. The index applies them as written and states every deviation.

  • Ph. Eur. 2.2.29 — Liquid chromatography
  • USP <621> — Chromatography
  • Ph. Eur. 2.2.46 — Chromatographic separation techniques (asymmetry, resolution, plate count)

Related methods

Peak integration practice

Drop-line versus valley-to-valley, tangent skim, baseline placement, and the reporting threshold — where a purity figure is actually decided.

Chromatographic purity

Gradient design for peptides

Why 25 minutes is the index minimum for the acylated incretins, and how gradient slope trades against run time and against the impurities you can see.

Chromatographic purity

System suitability

The injections that establish a chromatographic system was fit to produce the result, and the parameters the index checks before accepting a purity figure.

Quality system